PrimeLab 1.0 MULTITEST User Manual page 193

Photometer
Table of Contents

Advertisement

Nickel (HR)
(100)
Continued...
9) Unscrew the lid.
10) Add 5 drops "PL Nickel HR 3" liquid reagent to the test liquid in the
cuvette.
11) Replace the lid on the cuvette, swivel it carefully and insert it in the
sampling chamber of the PrimeLab. Align the arrow on the front of the
4
cuvette. Press TEST.
12) After the lapse of a 15 minute countdown the determined result is
displayed. Press button 4 to save the test result together with date and
4
time on the device.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
Notes
->
Ferrous Iron will interfere with this test and can influence the readings.
->
High EDTA levels (> 25 mg/l) will interfere with this test and can influence
the reading (low reading).
->
High cobalt levels (> 0.5 mg/l) will interfere with this test and can influence
the reading (high reading).
->
Polyphosphates in the sample does not influence the reading.
160
0 - 10 mg/l
Name on device: 100-Nickel-HR-liq
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
(Ni)
Liquid