Error Messages - PrimeLab 1.0 MULTITEST User Manual

Photometer
Table of Contents

Advertisement

Menu: SET (Setup)
Continued...
Calibration
The following refers to calibration of ID 111/113 (PTSA/Fluorescein):
For ID 111 and ID 113 it is essential to always use the same vial for calibration as well as
for testing as adapter shines from above and each vial has different bottom
Due to the above it is essential to always use the same vial for testing which has
been used for calibration!
If you select „PTSA-/ Fluorescein-adapter":
Select a 24mm glass vial for future PTSA and Fluorescein tests.
Fill vial with 10ml of distilled water. USE PIPETTE TO ENSURE 10ml!
Place vial into the PrimeLab. Do not close it!
Place PTSA-/Fluorescein-adapter on top of the open cuvette.
Press TEST.
4
Remove vial from PrimeLab.
Empty vial.
Fill vial with 10ml of reference-solution „500 ppb PTSA" / „100 ppb
Fluorescein". USE PIPETTE!
Place vial into the PrimeLab. Do not close it!
Place PTSA-/Turbidity-adapter on top of the open cuvette.
Press TEST. Calibation was sucessful!
4
Please use ONLY THIS VIAL for future PTSA/Flurescein tests!

Error Messages

During calibration and / or during the measurement, the following error
messages might be displayed on the PrimeLab:
Adapter not recognized - Possible causes:
Battery depleted
Ÿ
Adapter incorrectly (eg diagonally) placed
Ÿ
Wrong adapter used (eg PTSA instead Turbidity)
Ÿ
Adapter-lense (bottom) or optical path in the sample chamber
Ÿ
(PrimeLab) dirty or wet
Ÿ
Measurement failed - Possible causes:
Battery depleted
Ÿ
Adapter incorrectly (eg diagonally) placed
Ÿ
The identified water sample does not match the turbidity curve
Ÿ
Wrong adapter used (eg PTSA instead of Turbidity)
Ÿ
Adapter-lense (bottom) or optical path in the sample chamber
Ÿ
(PrimeLab) dirty or wet
Ÿ
"LED problem. Please check!" please abort the procedure and check that
chamber and especially the two "windows" on the sensor / LED level are
If the device fails to calibrate after a second attempt it must be returned to
SET 6
If the message reads "Calibration: ERROR!" or
there is no cuvette in the sampling chamber and that the sampling
clean (see also "Cleaning the device" on page A 3).
Restart the calibration procedure.
the manufacturer for inspection.