Measure Group - YOKOGAWA TA720 User Manual

Time interval analyzer communication interface
Hide thumbs Also See for TA720:
Table of Contents

Advertisement

4.11 MEASure Group

The commands in this group deal with measurement conditions. You can make the same settings and
inquiries as when the MODE key on the front panel is used.
:MEASure
:
:MEASure?
Function
Queries all settings related to the measurement
conditions.
Syntax
:MEASure?
Example
:MEASURE?
-> :MEASURE:MODE HHISTOGRAM;
FUNCTION PERIOD,A;SLOPE RISE
IM 704510-17E
;
FUNCtion
<Space>
?
MODE
<Space>
?
POLarity
<Space>
?
SLOPe
<Space>
?
?
PERiod
,
A
PWIDth
B
TI
AB
PPERiod
PTI
PWTI
PWPW
TSTamp
HHIStogram
ISI
POSitive
NEGative
BOTH
RISE
,
RISE
FALL
FALL
BOTH
:MEASure:FUNCtion
Function
Sets the measurement function or queries the
current setting.
Syntax
:MEASure:FUNCtion
{PERiod|PWIDth|TI|PPERiod|PTI|PWTI|
PWPW},{A|B|AB}
:MEASure:FUNCtion?
Example
:MEASURE:FUNCTION PERIOD,A
:MEASURE:FUNCTION?
-> :MEASURE:FUNCTION PERIOD,A
Description • "A|B" can be specified for the second
parameter when measuring period or pulse
width.
• Only AB can be specified for the second
parameter when measuring A-to-B time
interval (TI), period A & period B (PPERiod),
period A & A-to-B time interval (PTI), pulse
width A & A-to-B time interval (PWTI), pulse
width A & pulse width B (PWPW).
• "PWIDth|PWTI|PWPW" can be specified
when the sampling mode is set to inter-
symbol interference analysis.
4.11 MEASure Group
4-55
4

Advertisement

Table of Contents
loading