6.3

Setting the Trigger Delay

Procedure
VERTICAL
1
CH
2
CH
3
CH
4
CH
5
CH
6
CH
7
CH
8
CH
Explanation
Selectable Range of Delay
IM 701450-01E
TRIGGER
HORIZONTAL
TRIG D
SEARCH
ACTION
PRESET
ZOOM
MODE
V/DIV
TIME/DIV
SIMPLE
ENHANCED
DELAY
POSITION
LOGIC
SNAPSHOT
CLEAR TRACE
HELP
1.
Press SHIFT+POSITION(DELAY). The DELAY menu appears.
2.
Press the soft key corresponding to the desired time unit.
If you are not specifying a time unit (ms, us, ns, or ps), press the DELAY soft key.
3.
Turn the jog shuttle to set the delay.
Pressing RESET resets the value to 0 s.
Although the display usually shows the waveform before and after the trigger point, it is
possible to display the acquired waveform after a fixed time period elapses using the
trigger delay.
0 to 4 s (resolution is (1/sample rate)/10))
T (Trigger position)
Delay
Trigger point
Note
When T/div is changed, the delay remains unchanged.
If the time base is set to external clock, the trigger delay cannot be specified (delay is fixed 0
s).
• To exit the menu during operation, press ESC located
above the soft keys.
• For a description of the operation using a USB keyboard or
a USB mouse, see section 4.3.