5.9

Using the Linear Scaling Function

Procedure
VERTICAL
1
CH
2
CH
3
CH
4
CH
5
CH
6
CH
7
CH
8
CH
IM 701450-01E
TRIGGER
HORIZONTAL
TRIG D
SEARCH
ACTION
PRESET
ZOOM
MODE
V/DIV
TIME/DIV
SIMPLE
ENHANCED
DELAY
POSITION
LOGIC
SNAPSHOT
CLEAR TRACE
HELP
1.
Press one of the CH1 to CH8/4 keys to select the channel. The CH menu
appears.
You can select up to channel CH4 and CH8 on the DL7440 and DL7480, respectively.
2.
Press the Next (1/2) soft key
3.
Press the Linear Scale AX+B soft key to select ON.
4.
Press the A/B soft key to set the jog shuttle control to A.
5.
Turn the jog shuttle to set the A value.
6.
Likewise, set the B value.
7.
To attach a unit, press the Unit soft key to display the keyboard and enter the
unit name.
Note
Linear scaling is not available for the following waveforms.
Snapshot waveforms
Accumulated waveforms (however, linear scaling is possible on the accumulated
waveform acquired last.)
Linear scaling is set for each channel.
The scaling coefficient A and offset value B that you entered are held even if you turn OFF
the linear scaling function.
Computation is performed using the linear scaling results.
<For a description of this function, refer to page 2-4.>
• To exit the menu during operation, press ESC located
above the soft keys.
• For a description of the operation using a USB keyboard or
a USB mouse, see section 4.3.