Care

– Clean breathing valve and ventilation hoses each time
after use.
– Clean ventilator and medical gas hoses if heavily
soiled.

Stripping down

Remove Oxylog 2000 from holder.
1 Disconnect ventilation hose from socket.
2 Disconnect flow measuring hoses from sockets.
3 Unscrew medical gas hose from Oxylog 2000.
When disconnecting hoses, always grip the sleeve
and not the corrugations!
If this is not done, the corrugations or hose may be
torn from the sleeve.
4 Disconnect flow sensor from breathing valve – do
not twist or use force on the hose nozzles, as
this can damage the flow sensor.
5 Carefully detach flow measuring hoses from flow
sensor, pulling in the axial direction of the hose nozzles.
6 Detach angled connector from flow sensor.
Do not allow any objects to enter the flow sensor.
Do not purge with compressed air. The wind vane
inside may be damaged and cause measuring
errors!
Detach ventilation hose from breathing valve.
Stripping down
3
2
1
4
5
6
Care
17