Checking correct functioning
To be carried out before every use.
Verify completeness of Inhalette.
Check general state of Inhalette and of flowmeter
with gas connection.
Check tightness of screw connections of connec-
tion hose.
Make sure that the nozzle is fitted tightly onto the
atomizer.
Unscrew medication container, fill with medica-
tion and screw back into housing. Open metering
valve. A strong flow of mist is emitted from Inha-
lette.
Operation
F
Unscrew medication container, fill with medica-
tion.
Filling level 5 is equivalent to 5 mL,
Filling level 10 is equivalent to 10 mL.
Screw medication container back in place.
Inhalette with toggle switch valve (fixed flow):
Actuate toggle switch. A strong flow of mist is
emitted from Inhalette. To end nebulization actu-
ate the toggle switch again.
Instructions for use Inhalette
Inhalette with flowmeter (adjustable flow):
Open metering valve and set the desired size of
the aerosol particles using the metering knob,
see "Particle spectrum of the medication nebuliz-
er" on page 19.
Read flow at upper rim of ball.
To end nebulization close metering valve.
Shut down
Unscrew medication container, empty any resi-
dues, and rinse with distilled water.
Fill with about 10 mL of distilled water, screw
medication container back on, nebulize for about
1 minute to clean the atomizer nozzle.
Empty out any residual distilled water.
Disassembly
F
A
B
C
D
E
Unscrew connection hose (A).
Pull out mask or mouthpiece (B).
Unscrew medication container (E).
Pull atomizer (D) off.
Unscrew nozzle (C).
English
15