Care

Stripping down machine

Remove anaesthetic gas scavenging system
Stripping down machine
Swing up table top to remove piston pump:
(only possible in »standby«)
1 Swing lever as far as possible to unlock.
2 Hold breathing system by handle and lift out.
Do not damage flat seals – avoid contact with sharp
edges. It must be put down so that the sealing
elements are not subject to constant pressure!
3 Caution!
Temperature of hotplate for breathing
system equals approx. 60 °C.
4 Handle pointing up – lift out piston pump
– do not use force –
Do not damage flat seals – avoid contact with sharp
edges. It must be put down so that the sealing
elements are not subject to constant pressure!
1 Swing lever back to original position, otherwise table
top cannot be reclosed.
Remove anaesthetic gas scavenging system
Refer to separate Instructions for Use for anaesthetic gas
scavenging system.
Unplug anaesthetic gas scavenging hose from
anaesthetic gas scavenging system.
5 Press button to unlock hose connector and pull
downwards. Hose remains on connector.