Installing The Sensor; Installing Prosonic Flow P And Prosonic Flow 93W; (Dn 15 To 65 / ½ To 2½") - Endress+Hauser Proline Prosonic Flow 93 Brief Operating Instructions

Hide thumbs Also See for Proline Prosonic Flow 93:
Table of Contents

Advertisement

Proline Prosonic Flow 93
3.8

Installing the sensor

3.9

Installing Prosonic Flow P and Prosonic Flow 93W

(DN 15 to 65 / ½ to 2½")
3.9.1 Mounting the sensor
Prerequisites
• The installation distance (sensor distance) is known  ä 8.
• The sensor holder is already mounted  ä 14.
Material
The following material is needed for mounting:
• Sensor incl. adapter cable
• Connecting cable for connecting to the transmitter
• Coupling fluid for an acoustic connection between the sensor and pipe
Procedure
1.
Set the distance between the sensors as per the value determined for the sensor distance.
Press the sensor down slightly to move it.
Fig. 9: Setting the distance between the sensors as per the value for the sensor distance
a
Sensor distance
b
Contact surfaces of the sensor
2.
Coat the contact surfaces of the sensors with an even layer of coupling fluid
(approx. 0.5 to 1 mm / 0.02 to 0.04" thick).
3.
Fit the sensor housing on the holder.
Endress+Hauser
Installation
b
A0011529
19