Installing Prosonic Flow W Sensor; Installing Prosonic Flow W; (Dn 15 To 65 / ½ To 2½") - Endress+Hauser HART Proline Prosonic Flow 91 Operating Instructions Manual

Ultrasonic flow measuring system
Hide thumbs Also See for HART Proline Prosonic Flow 91:
Table of Contents

Advertisement

Proline Prosonic Flow 91
!
Endress+Hauser
3.6

Installing Prosonic Flow W sensor

3.6.1
Installing Prosonic Flow W (DN 15 to 65 / ½ to 2½")
Mounting the sensor
Prerequisites
• The installation distance (sensor distance) is known  ä 14.
• The sensor holder is already mounted  ä 15.
Material
The following material is needed for mounting:
• Sensor incl. adapter cable
• Connecting cable for connecting to the transmitter
• Coupling fluid for an acoustic connection between the sensor and pipe
Procedure
1.
Set the distance between the sensors as per the value determined for the sensor distance.
Press the sensor down slightly to move it.
Fig. 13:
Setting the distance between the sensors as per the value for the sensor distance
a
Sensor distance
b
Contact surfaces of the sensor
2.
Coat the contact surfaces of the sensors with an even layer of coupling fluid
(approx. 0.5 to 1 mm / 0.02 to 0.04") thick.
3.
Fit the sensor housing on the sensor holder.
Note!
• Avoid to use a thick layer of the coupling fluid (less is more).
• Clean and reapply new coupling fluid when sensor is removed from the pipe.
Fig. 14:
Fitting the sensor housing
Installation
b
A0011529
A0011527
19