4.4.2 Test Mode [MODE]

46
Trigger mode
When entering LIST MEAS for the first time, the system will
set Trig Mode to MAN and automatically performs a
measurement.
Trig Mode
INT
MAN
EXT
BUS
The [LIST MEAS] page will execute a list of up to 10
swept frequency or amplitude tests.
When [MODE] is set to SEQ and [TRIG] is set to
MAN, the [LIST MEAS] function will automatically
execute each test step on the list in sequence until the
last step is executed. The meter will then wait for the
trigger button to be depressed before repeating the
process.
When [MODE] is set to STEP and [TRIG] is set to
MAN, the [LIST MEAS] function will automatically
execute the first test step. The meter will then wait for
the trigger button to be depressed before executing the
next step. This process is repeated for each step on the
list.
Test Mode
Soft-key
Function
SEQ
Sequence measure all step by one
trigger.
STEP
Measure one step by one trigger.
Function
Internal Trigger. All ten list points
are swept continuous.
Manual Trigger. Each time the
instrument is triggered by [Trig]
key, the list points are swept one by
one.
External Trigger. Each time the
instrument is triggered by the
handler trigger pin, the list points are
swept one by one.
BUS Trigger. Each time the
instrument is triggered by SCPI
command, the list points are swept
one by one.
Measure Key