11.13.10 CORRection:SPOT:FREQuency

The CORRection:SPOT:FREQuency command sets the frequency point for
the specified frequency point correction.
Command
Syntax
Parameter
Example
Query Syntax
Query Response
Example

11.13.11 CORRection:SPOT:OPEN

This command executes the OPEN correction data measure for the specified
frequency correction.
Command
Syntax
Example

11.13.12 CORRection:SPOT:SHORt

This command executes the SHORT correction data measure for the
specified frequency correction.
Command
Syntax
Example
120
{on,off}
SEND> CORR:SPOT:STATe?<NL>
RET>
on<NL>
CORRection:SPOT:FREQuendy <value>
value
NR1,NR2 or NR3:Frequecny value.
A suffix multiplier can be used with this command. But
the unit "Hz" cannot be added.
SEND> CORR:SPOT:FREQ 100<NL>
SEND> CORR:SPOT:FREQ 10k<NL>
CORRection:SPOT:FREQuency?
<NR3>
SEND> CORR:SPOT:FREQ?<NL>
RET>
1.000000e+03<NL>
CORRection:SPOT:OPEN
SEND> CORR:SPOT:OPEN<NL>
RET>
pass/fail<NL> //Finish or Undone
CORRection:SPOT:SHORt
SEND> CORR:SPOT:SHOR<NL>
RET>
pass/fail<NL> //Finish or Undone