4.3 [LIST SETUP] Page

To specify frequency points and perform short
correction
Step 1.
Press the [MEAS SHORT] soft key, a
dialog message, "Short-circuit the test
terminals", will be displayed.
Step 2.
Connect the test fixture to the BNC terminals
and short-circuit the test terminals.
Step 3.
Press the [OK] soft-key. The LCR-6000
measures the short impedances (inductances
and resistance) at this frequency point.
During the measurement, a "SHORT
measurement in progress" dialog message will
be shown on the display.
When the measurement has finished,
"Correction finished" is displayed on the
screen. During the measurement, you can press
the [ABORT] soft-key to abort short
correction.
Press the [Measure] key and press the [LIST SETUP]
soft-key to open the [LIST SETUP] page. The List
Measure feature can perform automatic sweep
measurements by sweeping the frequency or signal level
through a maximum 10 list points.
Before using the List Measure feature, you have to
configure the list setup.
In the [LIST SETUP] page, you can configure each of
the following list measurement controls with the cursor
placed in the corresponding field.
Function mode [FUNC]
Test mode [MODE]
Parameter selection [FREQ[Hz], VOLT[V],
CURR[A]]
Limit parameter selection [LMT]
Lower and upper limits [LOWER] [UPPER]
LCR-6000 Series User Manual
41