Terminator; Hand Shake - Cubicon Lux User Manual

Table of Contents

Advertisement

6.1.5 TERMINATOR

6.1.6 Hand Shake

72
Soft-key
1200
9600
38400
57600
115200
The command terminator EOL uses the ASCII Line
Feed character by default. (LF, Decimal 10, Hex 0x0A, or
ASCII '\n')
The EOL of the command and the EOL of the
command return data are the same.
The terminator (EOL) can change the three types:
1. LF (Line feed, Hex 0x0A).
2. CR (Carriage Return, Hex 0x0D).
3. CR+LF.
The LCR-6000 PC software only supports LF types.
Soft-key
LF
CR
CR+LF
If handshaking is turned on, the meter will return the
commands it received from the PC first before returning
the result string of the command.
Example: PC sends idn?
The meter returns:
idn? LCR-6300 RevC1.0
To set up the Hand Shake feature
Step 1.
Press the [Measure] or [Setup] key.
Step 2.
Press the [SYSTEM] bottom soft-key.
Step 3.
Use the cursor key to select the
[HAND SHAKE] field.
Step 4.
Use the soft-keys to set the handshake setting.
Function
Sets the baud rate to 1200.
Sets the baud rate to 9600.
Sets the baud rate to 38400.
Sets the baud rate to 57600.
Sets the baud rate to 115200.
Function
Set up the EOL to LF(0Ah).
Set up the EOL to CR(0Dh).
Set up the EOL to CR+LF(0D 0Ah).