Figure ‎ 4 -2

4.2.1 Open Correction [OPEN]

Figure ‎ 4 -3
[OPEN/SHORT] Page
The LCR-6000 Series' OPEN correction capability
cancels errors due to the stray admittance (G, B) in
parallel with the device under test (Refer to Figure ‎ 4 -3).
Stray Admittance
The OPEN correction data is taken at all preset
frequency points, independent of the test frequency you
set. Refer to the tables on page 128 for the trimming
frequency point lists for each model.
To perform open correction
Step 1.
Press the [Measure] key.
Step 2.
Press the [OPEN SHORT] soft-key.
Step 3.
Use the cursor key to select the [OPEN] field.
Soft-key
ON
LCR-6000 Series User Manual
Function
Enables open correction.
37