5.2.8 Lower and Upper Limits

5.3 [BIN MEAS] Page

64
Soft-key
Function
1-BINS
Set to 1 bin
2-BINS
Set to 2 bins
3-BINS
Set to 3 bins
4-BINS
Set to 4 bins
5-BINS
Set to 5 bins
6-BINS
Set to 6 bins
7-BINS
Set to 7 bins
8-BINS
Set to 8 bins
9-BINS
Set to 9 bins
LCR-6000 Series built-in comparator can sort DUTs into
a maximum 10 levels (Bin1 through Bin9 and OUT)
using up to nine sets of primary parameter limits along
with one set of secondary parameter limits.
To enter the limit values for tolerance mode
Step 1.
Press the [Setup] key.
Step 2.
Press the [BIN SETUP] soft-key.
Step 3.
Use the cursor key to select [BIN 1 LOWER]
field
Step 4.
Enter the limit value using the entry keys.
When in the ABS and SEQ mode, please enter
the unit using the soft-keys. When in the PER
mode, the unit is %.
Step 5.
Repeat Step 4 using the cursor keys until you
have entered all limit values.
Step 6.
Use the cursor key to select [2nd LOWER]
field and enter the lower limit value for the
secondary parameter.
Step 7.
Select [2nd UPPER] field to enter the upper
limit value.
When the [Setup] key and the [BIN MEAS] soft-key is
pressed, the [BIN MEAS] page appears. The bin sorting
results are displayed in larger characters while the
measurement results are shown in normal characters.
Setup Key