4.1.2 Impedance Range [RANGE]

Table ‎ 4 -4
Impedance range mode
To achieve the best measurement result, it is
recommended to set the impedance range mode to
"Auto range". When using "Hold range", errors in
measurement results may occur if the measured value
exceeds the full scale of the selected range.
Impedance range mode
Mode
Function
overview
Auto
LCR-6000
Range
Series sets the
optimum
impedance
range for the
impedance of
the DUT
Hold
Measurement
Range
Nominal
LCR-6000
Range
Series sets the
optimum range
depending on
the nominal
value
Available impedance range
The hold range has nine impedance ranges: 10, 30, 100,
300, 1k, 3k, 10k, 30k and 100k.
The impedance range is selected according to the DUT's
impedance even if the measurement parameter is
capacitance or inductance.
LCR-6000 Series User Manual
Advantage
Disadvantage
You don't
The
need to select
measurement
range
time is longer
due to the
ranging time
No ranging
is performed with a fixed impedance ran
You need to
time is
select a proper
required
range
depending on
the value of
the DUT.
You don't
Valid only in
need to select
the sorting
range. No
mode
ranging time is
required
27