6.1.7 Error Code

Soft-key
Function
OFF
Turn off the Hand Shake feature.
ON
Turn on the Hand Shake feature.
If the error code setting is set to on, the meter will return
error codes if the wrong command or an invalid
command is received to help you to debug your control
program.
"*E00", //No error
"*E01", //"Bad command",
"*E02", //"Parameter error",
"*E03", //"Missing parameter",
"*E04", //"buffer overrun",
"*E05", //"Syntax error",
"*E06", //Invalid separator",
"*E07", //"Invalid multiplier",
"*E08", //"Numeric data error",
"*E09", //"Value too long",
"*E10", //"Invalid command"
"*E11", //"Unknown error"
To set up the Error Code feature
Step 1.
Press the [Measure] or [Setup] key.
Step 2.
Press the [SYSTEM] bottom soft-key.
Step 3.
Use the cursor key to select the
[ERROR CODE] field.
Step 4.
Use the soft-keys to set the error code setting.
Soft-key
OFF
ON
LCR-6000 Series User Manual
Function
Turn off the Error Code feature.
Turn on the Error Code feature.