4.2.2 Short Correction [SHORT]

Figure ‎ 4 -4
38
OFF
MEAS OPEN
DCR OPEN
Step 4.
Press the [MEAS OPEN] or [DCR OPEN]
soft-key. A dialog message, "Open-circuit the
test terminals" will appear.
Step 5.
Connect the test fixture to the BNC terminals.
Do not connect the DUT to the fixture.
Step 6.
Press the [OK] soft-key. The LCR-6000 will
then measured the open admittances
(capacitance and conductance) at the all test
frequency points.
During the measurement, an "LCR OPEN
measurement in progress" dialog message is
shown on the display.
When the measurement has finished,
"Correction finished" is displayed. During the
measurement, you can press the [ABORT]
soft-key to abort the OPEN correction.
Step 7.
After the OPEN measurement has finished,
the meter will enable OPEN correction
automatically. You can, however, still choose
to manually choose to enable or disable OPEN
correction by depressing the ON or OFF soft-
keys.
The short correction feature of the LCR-6000 Series
compensates for any residual impedances (R,X) that may
exist outside of calibration, such as the impedance of the
cables and the DUT connection points.
See Figure ‎ 4 -4 Residual Impedances.
Residual Impedances
To perform short correction
Disables open correction.
Starts open correction (AC).
Starts open correction (DC).
Measure Key