4.1.3 Test Frequency [FREQ]

Table ‎ 4 -7
Procedure for setting the impedance range [RANGE]
Step 1.
Press the [Measure] key.
Step 2.
Use the cursor key to select the [RANGE]
field.
Step 3.
Use the soft-keys to select the impedance
range mode or impedance range.
Soft-key
AUTO RANGE
HOLD RANGE
NOMINAL RANGE
INCR +
DECR -
LCR-6300: 10Hz~300kHz
LCR-6200: 10Hz~200kHz
LCR-6100: 10Hz~100kHz
LCR-6020: 10Hz~20kHz
LCR-6002: 10Hz~2kHz
Frequency range and resolution
Frequency range(F)
10.00Hz ≦ F ≦ 99.99Hz
100.0Hz ≦ F ≦ 999.9Hz
1.000kHz ≦ F ≦ 9.999kHz
10.00kHz ≦ F ≦ 99.99kHz
100.0kHz ≦F ≦ 300.0kHz
Frequency Accuracy: ±0.01% with 4-digit resolution
LCR-6000 Series User Manual
Function
Sets to AUTO range
Sets to HOLD range
Sets to NOMINAL range
Increments the impedance
range in the HOLD mode
Decrements the impedance
range in the HOLD mode
Resolution
0.01Hz
0.1Hz
1Hz
10Hz
100Hz
29