11.12 LIST Subsystem
The LIST or SWEEP Subsystem command group sets the List Sweep
measurement function, including the sweep point setting and limit values for
the limit function.
Figure 1 1-9
LIS T
11.12.1 LIST:PARAmeter
The LIST:PARAmeter command sets the list sweep parameter.
Command
Syntax
Parameter
Example
Query Syntax
Query Response
Example
11.12.2 LIST:STAT
The LIST:STAT command turns on/off the specified sweep point.
Command
Syntax
Parameter
Example
Query Syntax
LIST Subsystem Command Tree
:PARAmeter
:S TAT
:BAND
<n>,<value>,{A,B,OFF},<low>,<high>
LIST:PARAmeter {FREQ,VOLT,CURR}
Where, {FREQ,LEVEL} is:
FREQ
Sets the sweep parameter to frequency
VOLT
Sets the sweep parameter to voltage level
CURR
Sets the sweep parameter to current level
SEND> LIST:PARA VOLT<NL>
LIST:PARAmeter?
{FREQ,VOLT,CURR}
SEND> LIST:PARA?<NL>
RET>
FREQ<NL>
LIST:STAT <n>,{ON,OFF,1,0}
Where,<n> is:
n
NR1(1 to 10): List sweep point
ON or 1 Set this point to ON
OFF or 0 Set this point to OFF
SEND> LIST:STAT 1,ON<NL>
LIST:STAT? <n>
LCR-6000 Series User Manual
{FREQ,VOLT,CURR}
<n>,{ON,OFF}
115