To specify frequency points and perform open
correction
Step 1.
Press the [Measure] key.
Step 3.
Press the [OPEN SHORT] soft-key.
Step 3.
Use the cursor key to select the [SPOT] field.
Step 4.
Enter the frequency using the numeric entry
keys.
Soft-key
ON
OFF
CURRENT FREQ Use current frequency
MEAS OPEN
MEAS SHORT
Step 4.
Press the [MEAS OPEN] soft-key. A dialog
message, "Open-circuit the test terminals", will
appear.
Step 5.
Connect the test fixture to the BNC terminals.
Do not connect the DUT to the test fixture.
Step 6.
Press the [OK] soft-key. The LCR-6000
measures the open admittances (capacitance
and conductance) at this frequency point.
During the measurement, an "OPEN
measurement in progress" dialog message will
be shown on the display.
When the measurement has finished,
"Correction finished", will be displayed.
During the measurement, you can press the
[ABORT] soft-key to abort open correction.
Function
Enable this point.
Disables this point.
Starts open correction.
Starts short correction