Nominal Value For Tolerance Mode; Beep Feature; Total Number Of Bins [#-Bins] - Cubicon Lux User Manual

Table of Contents

Advertisement

5.2.5 Nominal value for tolerance mode

5.2.6 Beep feature

5.2.7 Total Number of Bins [#‐ BINS]
You must configure the nominal value when you use
tolerance mode as the limit mode for the primary
parameter.
In sequential mode the nominal value does not affect
sorting. In sequential mode, you do not need to configure
the nominal value.
To enter the nominal value
Step 1.
Press the [Setup] key.
Step 2.
Press the [BIN SETUP] soft-key.
Step 3.
Use the cursor key to select [NOM] field.
Step 4.
Enter the nominal value using the entry keys.
Use the soft-key to select the unit.
The beep feature behaves differently depending on
which beep mode is in effect.
To set up the beep feature
Step 1.
Press the [Setup] key.
Step 2.
Press the [BIN SETUP] soft-key.
Step 3.
Use the cursor key to select [BEEP] field.
Step 4.
Use the soft-keys to set beep feature.
Soft-key
Function
OFF
Turns the beep feature off
PASS
Beep when the comparator sorting
result is OK
FAIL
Beep when the comparator sorting
result is NG
LCR-6000 series specify nine bins (1-BINS to 9-BINS).
To choose total number of the bins
Step 1.
Press the [Setup] key.
Step 2.
Press the [BIN SETUP] soft-key.
Step 3.
Use the cursor key to select [#-BINS] field.
Step 4.
Use the soft-keys to select beep tone.
LCR-6000 Series User Manual
63

Advertisement

Table of Contents
loading

Table of Contents