5.1.2 Averaging Factor [AVG]

5.1.3 DC Bias Voltage [BIAS]

Soft-key
Function
30Ω
30Ω
50Ω
50Ω
100Ω
100Ω
The averaging feature allows you to obtain moving
average values of successive measurement results.
You can specify the averaging factor from integer 1 to
integer 256.
To set up the averaging factor
Step 1.
Press the [Setup] key.
Step 2.
Use the cursor key to select [AVG] field.
Step 3.
Use the soft-keys or number keys to enter
averaging factor.
Soft-key
INCR +
DECR -
This function will apply a DC bias voltage onto the DUT
while LCR meter is performing its AC measurement on
the DUT. The settable DC bias voltage range is from -
2.5V ~ 2.5V.
Procedure for setting DC bias voltage
Step 1.
Press the [Setup] key.
Step 2.
Use the cursor key to select [BIAS] field.
Step 3.
Use the soft-keys to select DC bias voltage.
Soft-key
Function
OFF
Turn off the DC bias voltage
2V
2V voltage is applied to the device under
test.
1.5V
1.5V voltage is applied to the device under
test.
-1.5V
-1.5V voltage is applied to the device under
LCR-6000 Series User Manual
Function
Increments the averaging factor in
steps of 1, 2, 4, 8, 16, 32,
64, 128 and 256.
Decrements the averaging factor in
steps of 1, 2, 4, 8, 16, 32,
64, 128 and 256.
53