2.2 Main Specifications and Features

2.2.1 Test Function

2.2.2 Equivalent Circuit

Table ‎ 2 -1

2.2.3 Ranging

2.2.4 Measurement Speed

Cs-Rs, Cs-D, Cp-Rp, Cp-D, Lp-Rp, Lp-Q, Ls-Rs, Ls-Q,
Rs-Q, Rp-Q, R-X, DCR, Z-θr, Z-θd, Z-D, Z-Q
Serial and Parallel
Equivalent Circuit
Circuit
Lp
L
Rp
Ls
Cp
C
Rp
Rs
Cs
Q=Xs/Rs, D=Rs/Xs, Xs=1/2πFCs=2πFLs
Auto, Hold and Nominal range. Total 9 Ranges.
About Nominal ranges: The LCR-6000 Series will
automatically select the best range according to the
nominal value.
Fast:
40 readings per second; 25ms/per reading
Medium: 10 readings per second; 100ms/per reading
Slow:
3 readings per second; 333ms/per reading
You can also specify the averaging factor within the range
of 1 to 256 in steps of 1 to improve performance.
LCR-6000 Series User Manual
Dissipation
Factor
D=2π
FLp/Rp=1/Q
D=Rs/2π
FLs=1/Q
Rs
D=1/2π
FCpRp=1/Q
D=2π
FCsRs=1/Q
Conversion
2
Ls=Lp/(1+D
)
2
2
Rs=RpD
/(1+D
)
2
Lp=(1+D
)Ls
2
2
Rp=(1+D
)Rs/D
2
Cs=(1+D
)Cp
2
2
Rs=RpD
/(1+D
)
2
Cp=Cs/(1+D
)
2
2
Rp=Rs(1+D
)/D
*1
*1
*1