2.3.2 Test Signal Level
2.3.3 Output impedance
2.4 Main Functions
2.4.1 Correction Function
2.4.2 Comparator Function (Sorting)
AC:
10.00mV- 2.00V (± 10%)
CV mode:10.00mV- 2.00V(± 6%)
100.0uA- 20.00mA (± 10%)
CC mode:100.0uA- 20.00mA(± 6%) @2VMax
DCR: ± 1V(2Vpp), Square wave, 3Hz up
0.033A(Max), Output impedance fixed 30Ω
30Ω, 50Ωand 100Ω
OPEN/SHORT correction:
Eliminates measurement errors due to stray parasitic
impedance in the test fixtures.
Bin sort
The primary parameter can be sorted into BIN1-BIN9,
AUX, OUT and HI/IN/LO for each of the primary
measurement parameters.
The sequential mode or tolerance mode can be selected as
the sorting mode.
Limit Setup
Absolute value, deviation value, and % deviation value
can be used for setup.
BIN count
Countable from 0 to 999999
LCR-6000 Series User Manual
17