4.1.5 Test Signal Voltage/Current Level [LEVEL]

Table ‎ 4 -13
32
The LCR-6000 Series' test signal voltage/current level
can be set as the effective value (RMS value) of a sine
wave of the test frequency from the unit's internal
oscillator.
When a constant voltage or current level measurement is
performed, an asterisk mark (*) appears at the end of the
LEVEL display.
Test signal level setting procedure
Step 1.
Press the [Measure] key
Step 2.
Use the cursor key to select [LEVEL] field
Step 3.
Use the soft-keys or numeric entry keys to
enter the test signal voltage/current level.
Soft-key
Function
INCR +
Refer to Table ‎ 4 -13
DECR -
ALC ON
Auto Level Control ON
ALC OFF
Auto Level Control OFF
Test voltage/current level that can be selected with
INCR+/DECR‐
INCR+/DECR-
10.00mV
100.0mV
1.00V
1.50V
INCR+/DECR-
100.0uA 500.0uA 1.00mA 5.00mA 10.00mA 20.00mA
Voltage range
10.00mV ≦ LEVEL ≦ 99.99mV
100.0mV ≦ LEVEL ≦ 999.9mV
1.00V ≦ LEVEL ≦ 2.00V
Current range
100.0uA ≦ LEVEL ≦ 999.9uA
1.00mA ≦ LEVEL ≦ 20.00mA
300.0mV
500.0mV
2.00V
Resolution
0.01mV
0.1mV
0.01V
Resolution
0.1uA
0.01mA
Measure Key