Nominal Value [Nominal] - Cubicon Lux User Manual

Table of Contents

Advertisement

NOTE

5.1.8 Nominal Value [NOMINAL]

56
In situations when the actual measuring Vac or
Iac goes beyond the extent that the ALC can
regulate, a warning message, "Failed! ALC
can't regulate!", will be shown at the bottom of
the screen to indicate to the operator that the
ALC can no longer regulate Vac/Iac.
An asterisk will be shown beside the LEVEL V
or A unit when the ALC is turned on.
Soft-key
OFF
ON
The Nominal Value parameter becomes displayed and
settable only when MON1 and MON2 are set in either Δ
or Δ% modes. The Nominal Value and the measured
main parameter will then be used to perform the Δ or
Δ% calculations.
Function
Turn off the Auto Level Control
Function
Turn on the Auto Level Control
Function
Setup Key