Installation

Requirements

Requirements
This section describes the requirements for the environment this instrument is used.

"Power Requirements"

"Operating Environment"

"Storaging/Shipping Environment"

"Installation Requirements"
"Power Cable"
Power Requirements
This instrument can operate from any single-phase AC power source supplying 100/
240 V (±10 %) at 50/60 Hz. The maximum power consumption is shown below.
Mainframe: 900 VA
Selector: 70 VA
Operating Environment
This instrument must be operated within the following environmental conditions:
+5 C to +40 C
Temperature:
Humidity:
20 % to 70 % RH, no condensation
Altitude:
0 to 2,000 m (6,561 ft.)
Storaging/Shipping Environment
This instrument must be stored/shipped within the following environmental
conditions:
-20 C to +60 C
Temperature:
Humidity:
10 % to 90 % RH, no condensation
Altitude:
0 to 4,600 m (15,092 ft.)
2- 4
Keysight B1507A User's Guide, Ed ition 1