Pulse Sweep

GUI Reference
I/V Measurement Software
You can change the settings of this terminal using the pull-down menu. For
details, see "Resource/Output Mode/Output Sequence" on page 4-50.
Number of Step
Specify the number of sweep steps from start to stop.
Compliance
Specify a current/voltage compliance value.
If the output exceeds this, the sweep measurement is terminated.
Power Compliance
Specify the power compliance value.
To enable this, mark the check box.
If the output exceeds this, the sweep measurement is terminated.
Not available for HVSMU.
Voltage Compliance
Specify a voltage compliance value for UHCU.
To enable this, mark the check box.
If the output exceeds this, the sweep measurement is terminated.
Available for UHCU.
The terminal outputs the staircase-swept pulse voltage/current.
Set the following parameters.
Sweep Mode
Select one from the following sweep modes.
LinearSingle (Sweep from start to stop in a linear incremental step)
LogSingle (Sweep from start to stop in a logarithmic incremental step)
LinearDouble (Sweep from start to stop to start in a linear incremental step)
LogDouble (Sweep from start to stop to start in a logarithmic incremental
step)
Start
Specify the sweep start value.
Stop
Specify the sweep stop value.