Bias
Pulse Bias
•
"Pulse Bias" on page 4-43
Outputs the voltage or current pulse that the peak output is constant.
•
"Step" on page 4-44
Outputs the constant voltage/current that the level is changed every time a sweep
measurement is finished.
•
"Pulse Step" on page 4-45
Outputs the voltage/current pulse that the peak level is changed every time a
sweep measurement is finished.
•
"Staircase Sweep" on page 4-47
Outputs the staircase-swept voltage/current.
•
"Pulse Sweep" on page 4-48
Outputs the staircase-swept pulse voltage/current.
The terminal outputs the constant voltage or current.
Set the following parameters.
•
Source
Specify an output voltage/current value.
The following parameters can be shown or hidden by clicking the chevron
Details.
•
Resource/Output Mode/Output Sequence
You can change the settings of this terminal using the pull-down menu. For
details, see "Resource/Output Mode/Output Sequence" on page 4-50.
•
Compliance
Specify a current/voltage compliance value.
If the output exceeds this, the sweep measurement is terminated.
•
Voltage Compliance
Specify a voltage compliance value for UHCU.
To enable this, mark the check box.
If the output exceeds this, the sweep measurement is terminated.
Available for UHCU.
The terminal outputs the voltage or current pulse that the peak output is constant.
Set the following parameters.
•
Source
Keysight B1507A User's Guide, Ed ition 1