Step 6. Specify the drive conditions that are not extracted from the loaded measurement
data.
1. Click the Drive Conditions tab at the left of work area to display the Drive
Conditions area.
2. Specify the following drive conditions that are not extracted from the
measurement data.
Switching
Frequency
Rg(ext)
Irms, D
Step 7. Click the Start button
Keysight B1507A User's Guide, Ed ition 1
Sweep stop value
Sweep mode
Log (logarithmic sweep) or Linear (linear sweep)
Number of sweep steps
Switching frequency
In this example, this parameter already specified by Step.4.
Gate input resistance in the external drive circuit
Irms: Effective drain current (the root mean square of the
load current)
D: Duty cycle.
Specify Irms or D. 
The relation between Irms and D is Irms = Id  D.
of toolbar to start the power loss calculation.
How To Perform Measurement
How To Calculate Power Loss
3- 23