Keysight Technologies B1507A User Manual page 70

Power device capacitance analyzer
Table of Contents

Advertisement

Table 2-2
Table 2-3
Installation
Inspection and Installation
4. Connect the System cable between the selector and the test fixture.
System Cable Connections between Mainframe and Selector
Mainframe
Interlock
GNDU
SMU1
Sense
Force
SMU2
Force
MFCMU
Digital I/O
System Cable Connections between Selector and Test Fixture
Selector
AC/DC Guard
Emitter/Source
Base/Gate
Collector/Drain
Earth terminal
Interlock
2- 12
Label of Cable
System Cable 1
none
none
GNDU
GNDU
1S
1S
1F
1F
2
2
CMU Cable
Hpot
HP
Lpot
LP
Hcur
HC
Lcur
LC
Digital I/O Cable
none
none
Label of Cable
System Cable 2
T4
T4
T3
T3
T1
T1
T2
T2
none
none
none
none
Keysight B1507A User's Guide, Ed ition 1
Selector
Interlock
GNDU
Sense (1S)
SMU
SMU1
Force (1F)
Force (2F)
HVSMU
SMU2
Hpot (HP)
MFCMU
Lpot (LP)
Hcur (HC)
Lcur (LC)
Digital I/O
Test Fixture
AC/DC Guard
Emitter/Source
Base/Gate
Collector/Drain
Earth terminal
Interlock

Advertisement

Table of Contents
loading

Table of Contents