GUI Reference
I/V Measurement Software
Set the following parameters.
Step Mode
Select one from the following level change operations.
LinearSingle: Peak level is changed by a constant step value.
ListSingle: Peak level is changed according to the listed levels.
Start
Specify the first pulse peak value.
Available for the "LinearSingle" Step Mode.
Stop
Specify the last pulse peak value.
Available for the "LinearSingle" Step Mode.
List
Create the list of the output levels.
Available for the "ListSingle" Step Mode.
To insert a new step entry before the selected step value, click the Insert
button
.
To move the selected step value up/down, click the Move button
To delete the selected step value, click the Delete button
The following parameters can be shown or hidden by clicking the chevron
Details.
Resource/Output Mode/Output Sequence
You can change the settings of this terminal using the pull-down menu. For
details, see "Resource/Output Mode/Output Sequence" on page 4-50.
Base
Specify a pulse base voltage/current value.
Number of Step
Specify the number of steps.
Available for the "LinearSingle" Step Mode.
Compliance
Specify a current/voltage compliance value.
If the output exceeds this, the sweep measurement is terminated.