Low Current Semiconductor Measurements
Using the B2980A Series Ammeter
Keysight B2981A/B2983A Femto/Picoammeter
Keysight B2985A/B2987A Electrometer/High Resistance Meter
Technical Overview

Advertisement

loading

Summary of Contents for Keysight Technologies B2981A

  • Page 1 Keysight Technologies Low Current Semiconductor Measurements Using the B2980A Series Ammeter Keysight B2981A/B2983A Femto/Picoammeter Keysight B2985A/B2987A Electrometer/High Resistance Meter Technical Overview...
  • Page 2 4. Integrated capability to analyze measurement data via histograms, time domain graphs, etc. The Keysight B2981A/B2983A Femto/Picoammeter and B2985A/B2987A Elec- trometer/High Resistance Meter provide both superior performance of 10 aA minimum resolution and improved measurement conidence by supporting all of the features listed above.
  • Page 3 03 | Keysight | Low Current Semiconductor Measurements Using the B2980A Series Ammeter - Technical Overview Measurement Basics and the B2980A Series Performance 1. Low noise, low impedance and low current ammeter provides stable measurement data The B2980A Series ammeter has the following features: –...
  • Page 4 04 | Keysight | Low Current Semiconductor Measurements Using the B2980A Series Ammeter - Technical Overview 2. Floating ammeter design provides stable measurements for both loating and grounded DUTs (continued) Grounded DUT coniguration Figure 2 shows the case where one of the ends of the DUT is grounded. In this case the con- nection scheme shown in Figure 1 cannot be used.
  • Page 5 05 | Keysight | Low Current Semiconductor Measurements Using the B2980A Series Ammeter - Technical Overview 3. Measurement speed, iltering/averaging and zero cancelation functions improve measurement data quality (continued) Null (Zero cancel) capability: To remove the effects of any residual offset currents, you can have the instrument automatically subtract out these currents by pressing the Null key.
  • Page 6 B2980A Series steps shown in this example (although the ammeter setup is identical). N1254A-101 Ammeter Triax cable Of course, when using the B2981A/B2983A you need to supply an external Coax adapter cable voltage source to measure the I/V characteristics. adapter...
  • Page 7 View function key. Note: In case of the B2981A/83A the instrument boots-up into this state and these operations are not necessary. Step 2. Check if the Voltage Source ield is set to 0 V. If not then follow the steps shown below to set the output voltage to zero volts.
  • Page 8 Step 6. (1) Continue rotating the [Knob] and increasing the voltage until the current is between 10 to 20 mA. Note: If using the B2981A/83A then increase the external voltage source voltage. (2) The LED starts to emit light at around 1.5 V.
  • Page 9 This example outlines how to plot LED current versus voltage (I/V) characteristics and it only applies to the B2985A/87A. Note: If using the B2981A/83A then you need to plot the current measured by the B2981A/83A and the voltage applied by the external voltage source on a PC to make this same type of I/V plot.
  • Page 10 10 | Keysight | Low Current Semiconductor Measurements Using the B2980A Series Ammeter - Technical Overview Example 2. LED forward diode I/V characteristics (continued) Step 3. (1) The Source function graphic changes to linear single sweep (upper right corner of screen). (2) Edit the sweep parameters as shown in the igure to the following values: - Start: 0 V - Stop: 1.8 V (or the anode voltage that lows about 10 ~ 20 mA anode current.)
  • Page 11 11 | Keysight | Low Current Semiconductor Measurements Using the B2980A Series Ammeter - Technical Overview Example 2. LED forward diode I/V characteristics (continued) Step 5. Press the [Single] key. The Trigger indicator changes to [ARM], and the sweep measurement begins. By pressing the [Auto Scale] function key, intermediate measurement data can be seen on the 10 X-Y graph.
  • Page 12 (BP-07-10-14) Keysight B2900 Precision Instrument Family The B2900Family lines up products for both precision source and precision measurement. www.keysight.com/ind/b2900a This information is subject to change without notice. © Keysight Technologies, 2014 Published in USA, September 4, 2014 5991-4952EN www.keysight.com...