Preparing Measurement

Set the instrument output off before connecting or disconnecting connection
wire.
Press the mainframe front panel Stop key to set the source output off. And
confirm that the mainframe front panel High Voltage indicator is not lit.
To prevent electrical shock and DUT damage, do not connect or disconnect the
DUT while the instrument is applying voltage or current.
When you touch the DUT after measurement, devise a countermeasure of
residual charge and heat to prevent electrical shock and burn. Use glove and
any tool. Also have enough time for discharge and radiation.
(DUT).
For more information on the test fixture and its accessories, see the following
sections:
"To Connect Selector and Test Fixture" on page 2-11
"3-pin Inline Package Socket Module" on page 2-17
"Accessories for Connecting a DUT" on page 2-18
that is adequate for your DUT.
For more information, see the following sections:
"To Connect Selector and Test Fixture" on page 2-11
"3-pin Inline Package Socket Module" on page 2-17
"Accessories for Connecting a DUT" on page 2-18
If your DUT is diode, connect a cathode terminal of your DUT to the
Collector/Drain terminal of test fixture and a anode terminal of your DUT to the
Emitter/Source terminal of test fixture.
3- 10