How To Perform Measurement
How To Perform I/V Measurement
LogDouble
Start、Stop
Enter the sweep start value in the Start and the sweep stop value in the Stop.
Voltage Compliance
Enter the voltage compliance value in the Voltage Compliance.
If the output voltage of measurement unit connected to a drain terminal is more
than this value, a measurement unit stops the output.
By clicking the
For more information on setting items, see "Setup" on page 4-38.
Step 5. Enter the setup conditions for the pulse output in General settings at the left side of
Setup area.
Auto Period
Pulse Period
3- 14
Logarithmic-sweep from the Start (sweep start value) to the
Stop (sweep stop value) and then return logarithmic-sweep
from the Stop to the Start。
chevron, the area expands to set more detail conditions.
Details
Automatic adjustment for the pulse period.
To enable this function, mark the check box.
Pulse period.
However, if the Auto Period function is enabled, the pulse
period is automatically adjusted to a minimum period
corresponding to the pulse width.
Keysight B1507A User's Guide, Ed ition 1