Chapter 10
| Initial start up
|
Dionex IC Pure system
58

Putting the system into operation

Note Allow the system to warm up or to cool down to room temperature before starting it up for
the first time.
CAUTION Check that all tubing connections have been made as specified in the "Connections of
the Dionex IC Pure system" on page 22 section.
Switch the system on by pressing the "On/Off" button. After a compulsory rinse, the system returns to
the last operating mode (Interval or Nonstop mode with IC system in Analyzer mode).
Note If an IC System is connected to the unit and switched on, "Analyser" will appear on the
control display of the Dionex IC Pure system. The Nonstop and Interval mode are then no longer
available. The user menu can continue to be used, however.
Note To vent air from the system, switch the system 3 times successively to "Rinsing" in the menu
and discard approx. 5 L of water each time with the manual dispenser. The ultra pure water limiting
value may come on during this process.
Use this 'NONSTOP'-key to switch the system to the "Nonstop" operating mode.
When the unit has successfully reached the pure water value specified, return the unit to the
"Interval" mode automatically or by pressing the "INTERVAL" button while it is in the "Nonstop"
mode.