Initial Operation; Reservoir Filling And Cartridge Rinse Up - Thermo Scientific EASYpure II RF Operating Manual And Parts List

Series 1305. reservoir feed water purification system
Table of Contents

Advertisement

Reservoir Filling and
Cartridge Rinse Up
Thermo Scientific

Initial Operation

Section 7
Note Cartridge rinse up procedure must be followed after each cartridge
and/or filter replacement.
Note In the event that the feedwater reservoir is overfilled, allow the excess
water to drain from the reservoir through the overflow drain tubing before
replacing the Ventgard cap. This will prevent wetting the filter element.
Warning Use a properly grounded electrical outlet of correct voltage and
current handling capacity.
Warning This device is to be used with water feeds only. Sanitizing/
cleaning agents must be used in compliance with instructions in this
manual. Failure to comply could result in explosion and personal injury.
1. Remove the molded plastic Ventgard cap.
2. Carefully pour pretreated feedwater into the reservoir until the water is
level with the top of the sight gauge on the front of the unit (see
Figure 7-1). Replace the Ventgard cap.
3. The power connection is located on the right side of the unit, in the
upper back corner.
4. Determine which power cord you need (this will be based on your
country and outlets available in your lab). A 120V power cord with (2)
1.6 amp fuses and 240V power cords with (2) 0.63 amp fuses are
provided with the unit.
5. Remove the fuse holder, install the fuses included with the power cord
to be used, and reinstall holder.
6. Verify power switch is turned off and attach receptacle end of power
cord into the power socket.
7. Plug other end of power cord into facility power. Turn main power on
at power entry module.
s
s
s
s
EASYpure II RF
7-1