Thermo Scientific dionex ic pure water Operating Instructions Manual page 39

Table of Contents

Advertisement

Step
Action
1
a. Remove the tube jumper from the
b. Attach the "connecting hose 20 L tank to IC Pure"
.
Note If you are not using a pretreatment cartridge,
skip to step d.
c. If you are using a pretreatment cartridge in the unit,
d. Attach the tube from the 'IN' nozzle on the unit to one
Note Either of the two connections on the external 20
L tank may be used to attach the tube to the unit.
e. Attach the o.d. 8 mm (0.31 in) tube into the right
Thermo Scientific
on the right side of the Dionex IC Pure system.
supplied with the unit to the
the „"Drain hose IC Pure (d8 mm (31 in) with quick
disconnect)" to
"OUT"
connection on the unit. Route
the tube from the
"OUT"
nozzle with a downward
slope to the drain
remove the tube jumper from the left concentrate
connection on the Dionex IC Pure system and route a
o.d. 8 mm (0.31 in) tube to the left connection of the
20 L tank.
of the two connections on the external 20 L tank.
connection of the external 20 L tank for tank overflow
and route the tube with a downward slope to the
drain.
Figure
"IN"
connections
"IN"
connection. Attach
IN / OUT Connections
o.d. 8 mm (0.31
in) tube
a.)
c.)
Concentrate
connection
o.d 8 mm (0.31 in) tube coming from
concentrate connection of Dionex IC Pure
system. Connected only if pretreatment
d.)
cartridge is used with IC Pure system.
20 L tank
connections
e.)
20 L tank
connection for
tank overflow
Installation |
Chapter 7
b.)
o.d. 8 mm (0.31 in) tubes
Dionex IC Pure system | 29