Chapter 13
| Maintenance
Step
Action
1
Switch off the unit and close the supply of feedwater if your unit
is connected with an external tank filling station.
2
a. When using a standard unit with an internal 5 L tank,
b. When using an external 20 L tank, take the Take the "drain
Note Let the remaining contents of both tanks drain
completely
.
3
a. Once the remaining contents of both tanks have been
b. Switch the system on.
|
Dionex IC Pure system
96
disconnect the tube jumper at the 'Internal tank inlet'
connection. Take the "drain piece for tank IC Pure" (o.d. 8
mm (0.31 in) tube with female quick disconnect) tube
supplied with the unit and attach it to the 'Internal tank
outlet' connection for the tube jumper and route the tube
with a slight downward slope to the drain (see
" Installation" on page 21). When you hear an audible click,
the quick-connectors have been inserted correctly.
piece for tank IC Pure" (o.d. 8 mm (0.31 in) tube with
female quick disconnect) supplied with it and attach it to
the bottom connection on the front of the 20 L tank and
route the tube with a slight downward slope to the drain
(see "Illustration of drain" on page 33). When you hear an
audible click, the quick-connectors have been inserted
correctly.
drained completely, remove the draining tubes from the
unit or from the external 20 L tank, reattach the tube jumper
to the Dionex IC Pure system and manually refill the
internal 5 L or external 20 L tank completely.
Figure
Internal tank outlet unit
a.)
Tube jumper
Internal tank outlet unit
o.d. 8 mm
(0.31 in) tube
b.)
Tube for draining
20 L tank
connection
Tube to unit
Thermo Scientific