10. Putting the system into operation

Note:
The tank lid must be tightly closed for the sterile overflow to function properly.
18
TKA-Pacific
NONSTOP
enter
UV
menu
INTERVAL
1. Connect the raw water inlet of your pure water system to
the pretreatment and this to a water tap that can be shut
off. The connector (R ¾") of the pure water system is
marked „Raw water".
2. Connect the concentrate outlet to the on-site drain. The
connector (R ¾") of the pure water system is marked "Con-
centrate". The drain to the stewer must be max. Are 1 m
above the rinsing water connector of the unit.
TAKE CARE! The concentrate must be able to pass to
drain under free gravity fall!
3. Pacific-RO with external tank (optional):
Use the Ø 8 mm hose provided to connect the pure water
outlet of the system to the pure water inlet of the tank.
4. Connect the cable of the level switch to the 4-pin plug on
the system.
Optional: Sterile overflow (accessory).
Plug the hose for the sterile overflow, Ø 8 mm, in the over-
flow at the back of the tank and connect this to drain.
Thermo Scientific Barnstead Pacific RO
Pure water system