Autoclaving The 0.2 Μm Final Filter - Thermo Scientific dionex ic pure water Operating Instructions Manual

Table of Contents

Advertisement

Step
Action
1
Rotate the used sterile filter in a clockwise direc-
tion to remove it.
2
Use an autoclave to sterilize the sterile filter in a
steam autoclave at 121 °C for 30 minutes. You can
repeat the sterilization process for the sterile filter
up to 5 times.
When sterilization is completed, screw the sterile
filter back into the 1/4" thread on the sampling tap
(see "Replacing the 0.2 µm final filter" on
page 92).
Thermo Scientific
Autoclaving the 0.2 μm final filter
Note The lifetime of the sterile filter can be prolonged by sterilizing it. Proceed as follows to
autoclave the sterile filter.
Figure
Note If you wish to remove purified water and the flow
rate is too low, this is an indication that the sterile filter
is clogged. In this case, read the description in section
"Troubleshooting" on page 99, or replace the sterile
filter with a new one.
Maintenance |
Chapter 13
1/4" thread
0.2 µm final filter
Dionex IC Pure system | 93