Sample Chromeleon Cds Software Settings For Dionex Ics-5000 Dionex Ics-5000 And Dionex Ics-3000 System; Connecting An Additional Ion Chromatograph Dionex Integrion And Dionex Ics-2100, Dionex Ics-4000, Dionex Ics-5000+, Dionex Ics-5000 Or Dionex Ics-3000 System - Thermo Scientific dionex ic pure water Operating Instructions Manual

Table of Contents

Advertisement

Chapter 7
|
Installation
Connecting an additional ion chromatograph Dionex Integrion and Dionex ICS-2100, Dionex
ICS-4000, Dionex ICS-5000
Step
Action
1
Switch off the Dionex IC Pure system and stop pumping of pure water
for the IC system using the Chromeleon CDS software.
2
Take the T piece and unscrew all three union nuts.
|
Dionex IC Pure system
42
Sample Chromeleon CDS Software settings for Dionex ICS-5000
Dionex ICS-5000 and Dionex ICS-3000 system
+
, Dionex ICS-5000 or Dionex ICS-3000 system
Note If you install a T piece you can also connect a second Dionex Integrion and Dionex
ICS-2100, Dionex ICS-4000, Dionex ICS-5000
the feed line to the first IC system. Both IC systems will then continue to be controlled using the
Chromeleon CDS software. The IC system that is connected to the control line at the unit serves as
the master. The second IC system does not have to be connected to the Dionex IC Pure system. If the
IC system connected to the Dionex IC Pure system is switched off, neither of the IC systems will
pump pure water from the Dionex IC Pure system.