Chapter 7
|
Installation
Step
Action
3
Take the PFA tube 1/8" supplied with the Dionex IC Pure system and cut
off the required length what you need.
As illustrated in the diagram, slide the union nut with the cutting ring
and taper coupling over the 1/8" tube.
4
a. Insert the 1/8" tube with the cutting ring and taper coupling
b. Press the reducing adapter into the "to suppressor"
5
Take the precision metering valve and plug in the 1/8" tube coming
from the Dionex IC Pure system into the left side from the precision
metering valve.
6
Proceed the same way as describe in step 5 with the 1/8" tube into the
right connection side of the precision metering valve. This tube is
connecting to the „"Regen In" connection of the suppressor from the
Dionex Integrion and Dionex ICS-2100, Dionex ICS-4000, Dionex
ICS-5000
|
Dionex IC Pure system
44
into the reducing adapter and tighten the union nut on the
fitting.
connection on the Dionex IC Pure system. When you hear an
audible click you can be sure that the quick-action-fasteners
have been inserted correctly.
+
, Dionex ICS-5000 or Dionex ICS-3000 system.
Figure
Cutting ring
1/8" tube
Union nut
Taper coupling
Reducing adapter
1/8" tube
b.
Device Connection
Reducing adapter
Suppresor
Left side
Right side
Precision metering valve
1/8" tube
see "Schematic illustration with connected
suppressor"
Union nut
1/8" tube
Thermo Scientific