Chapter 7
|
Installation
Step
Action
Connecting water tubings from IC Pure system to an ion chromatograph
Dionex Integrion and Dionex ICS-2100, Dionex ICS-4000, Dionex ICS-
5000+, Dionex ICS-5000 or Dionex ICS-3000 system
Note If hooking up more than one ion chromatograph, skip
steps 3-6 below. Following tubing installation "" on page 42.
Note If connection the suppressor to the IC Pure unit, skip
steps 3-6 below. Following tubing installation on "Connecting
the Suppressor to the Dionex IC Pure system" on page 43.
3
Take the reducing adapter supplied with the Dionex IC Pure sys-
tem and carefully unscrew the union nut from the fitting.
Note Ensure that the cutting ring and taper coupling that fit
inside the union nut are not lost when you unscrew the union
nut.
4
As illustrated in the diagram, slide the union nut with the cutting
ring and taper coupling over the 1/8" tube for the pump from the
Dionex Integrion and Dionex ICS-2100, Dionex ICS-4000, Dionex
ICS-5000
5
a. Insert the 1/8" tube with the cutting ring and taper coupling
b. The other end of the 1/8" tube is connecting to the "Degas in"
|
Dionex IC Pure system
38
+
, Dionex ICS-5000 or Dionex ICS-3000 system.
into the reducing adapter and tighten the union nut on the
fitting.
connection from the Degas module on the Dionex Integrion
and Dionex ICS-2100, Dionex ICS-4000, Dionex ICS-5000
Dionex ICS-5000 or Dionex ICS-3000 system. (see
"Schematic illustration with connected suppressor" )
Figure
Reducing adapter
Cutting ring
Union nut
Taper coupling
Reducing adapter
+
,
Union nut
1/8" tube
1/8" tube
Union nut
Thermo Scientific