System control |
Chapter 12
Note Use Menu button to scroll through menu items. Use enter only to save changed values.

Set the limiting value for temperature

The maximum operating temperature limit for the system is set here. Should this temperature be
exceeded, the error message "Max. temperature" is triggered. This is shown in the 3rd line of the
display.
Basic setting: 35 °C
Setting range: 1 - 50 °C
After entering the OEM menu, press the "Menu" button one time. The display shows:

Set the rinsing time

In this Menu point you can set the manually rinsing time. Rinse time is the amount of time the V4
Solenoid valve is opened to flush water to drain. This is helpful to purge air or impurities from the
system.
Basic setting: 30 sec.
Setting range: 10 - 60 sec.
After entering the OEM menu, press the "Menu" button two times. The display shows:

Change the disinfection time

The disinfection time can be set in this menu. Disinfection time is the length of time required by the
instrument to circulate the cleaning solution in the system.
Basic setting: 30 minutes
Setting range: 15 - 90 minutes
After entering the OEM menu, press the "Menu" button three times. The display shows:
Thermo Scientific
Dionex IC Pure system | 73