Chapter 13
| Maintenance
|
Dionex IC Pure system
82

Disinfection

The Dionex IC Pure system can only be disinfected as a 'stand-alone' process and cannot be
performed while in operation with an IC system.
Perform the following steps if an ion chromatograph system is connected to the Dionex IC Pure
system:
Switch off the IC system and remove the level switch cable for the Dionex Integrion and Dionex
ICS-2100, Dionex ICS-4000, Dionex ICS-5000
from the Dionex IC Pure system.
Disconnect the tubes for the IC system from the connections at the Dionex IC Pure system.
Required for disinfection is the disinfection adapter (item no: 50133431) purchase separately.
Use the following disinfecting agent:
Cleaning Solution, 1 syringe, item no.: CMX 25.
Note To ensure effective disinfection of your unit fill the internal 5 L or external 20 L tank 80%
with water. You can read off the level of both tanks using the additional tank indicator on the unit.
Always wear protective gloves when handling the cleaning solution.
Always wear safety goggles when handling the cleaning solution.
CAUTION If your skin should come into contact with a chlorine product, immediately rinse it
with ample, fresh water. If your eyes come into contact with the disinfecting agent, immediately rinse
them with ample, fresh water and immediately contact a physician.
+
, Dionex ICS-5000 or Dionex ICS-3000 system