Illustration of drain siphon for inserting the rinsing water and
concentrate outlet tube.
Note
Shorten the rinsing water and concentrate outlet tube to
the required length and route it to an atmospherically
vented drain.
The tubes that run from the Dionex IC Pure system to
the drain must be routed with a downward slope and
without any kinks or restrictions to ensure proper
drainage.
If a standard drain siphon is in place, the ends of the
tubes must be located at least 20 mm (0.75 in) above
the drain. Attach the tubes in such a manner that they
remain in their position.
Thermo Scientific

Illustration of drain

Overflow tube
Rinse water tubing
Minimum distance
20 mm/ 0.75 inch
Dionex IC Pure system | 33
Installation |
Chapter 7
Drain