System control |
Chapter 12

Set the sending interval

In this menu the sending interval between transmissions of measured values and error messages to
the RS 232 is set.
Basic setting: 1 hour
Setting range: 0.5 - 12 hours
After entering the OEM menu, press the "Menu" button seven times. The display shows:

Language selection

Basic setting: English
Setting range: Deutsch, English, French
After entering the OEM menu, press the "Menu" button eight times. The display shows:

Select units, conductivity/resistance

Basic setting: Resistance M
cm
·
Setting range: Resistance M
cm, Conductivity µS/cm
·
After entering the OEM menu, press the "Menu" button nine times. The display shows:
Thermo Scientific
Dionex IC Pure system | 75